Updated fields: d9, g3, g6, h2, h3, h4, h6, h10.3539700 padgett model s slimline electric dermatome was returned for evaluation.Device history record (dhr) review - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - received hand piece s-2019 with the 2¿ width clip, and also received hand piece s-2015 from the same end-user.S-2015 stopped functioning but was not involved in an injury incident.Dermatome cord was received with the hp, no width clips.Investigation completed for both units.No visible damage on the head, minor wear mark from the blade on the oscillating pin and eccentric screw.Both hand pieces were received with the hex nuts that attach the head assembly to the handle assembly installed backwards.Pic: ¿s-2015 head + handle¿ and ¿s-2019 head + handle¿ based on the picture included with the complaint the nuts were attached correctly during the injury incident." root cause - width clip damage could not be eliminated as a possible root cause.The burr, center clip, would have contacted the skin after the blade cut along the leading edge.The impact that caused the two cap side nicks could have created the gap in flatness spec, neither surface contacts the head assembly or blade.The bottom side of the clip shows the blade was contacting the clip as it oscillated, the friction was enough to mark the clip (blackish color is a result of friction from the blade reacting to the anodizing layer on the clip).As we have seen with s-2015 the moisture damage caused the motor to seize.While s-2019 functions now it¿s believed the moisture-induced damage will eventually seize this motor as well.Any loss of power during the procedure would certainly contribute to the failure.
|