MPRI ATTAIN STABILITY QUAD MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P Back to Search Results
Model Number 479888
Device Problems Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cardiac Tamponade (2226); Cardiac Perforation (2513)
Event Date 03/30/2022
Event Type  Injury  
Event Description
It was reported that during implant of the left ventricular (lv) lead, there was a failure in slitting the guide catheter after lead placement.The lv lead was confirmed to have been dislodged from a considerably pulled position under fluoroscopy.Afterwards, an attempt was made to place the lv lead again, but a patient abnormality was noticed after guide catheter cannulation.As blood pressure did not touch and the patient was not conscious, an echocardiogram was performed and confirmed a cardiac tamponade had occurred due to lv lead perforation.A pericardial puncture and cardiac massage were performed.Then, percutaneous cardiopulmonary support and an intra-aortic balloon pump (iabp) were inserted, and a thoracotomy was performed.Hemostasis was successfully achieved.Post evaluation of the patient revealed that the coronary sinus region that was fixated with the lv lead screw, was reported to have ruptured.The lv lead was not implanted.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN STABILITY QUAD MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14160122
MDR Text Key289615794
Report Number2649622-2022-07546
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/27/2023
Device Model Number479888
Device Catalogue Number479888
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2022
Date Device Manufactured01/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention; Hospitalization;
Patient Age79 YR
Patient SexFemale
-
-