Model Number 3758
Device Problems Vibration (1674); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Product return was requested but not received so far.Full evaluation will occur upon receipt of returned product.
 
Event Description
Brush head ¿jumps¿ off the magnet mid-brushing and the toothbrush/brush head then vibrates very strongly - oral-b [device connection issue].Case narrative: male consumer via e-mail stated that the oral-b toothbrush head ¿jumped¿ off of the magnet mid-brushing with their oral-b io series 9n toothbrush and the oral-b toothbrush/toothbrush head then vibrated very strongly, making brushing unpleasant.No injury was reported.
 
Event Description
Brush head ¿jumps¿ off the magnet mid-brushing and the toothbrush/brush head then vibrates very strongly - oral-b [device connection issue] case narrative: male consumer via e-mail stated that the oral-b toothbrush head ¿jumped¿ off of the magnet mid-brushing with their oral-b io series 9n toothbrush and the oral-b toothbrush/toothbrush head then vibrated very strongly, making brushing unpleasant.No injury was reported.(b)(6)2022 case update: the suspect product lot was ah04010059.No injury was reported.
 
Manufacturer Narrative
(b)(6)2022 product investigation results: product return was received and investigated.Product investigation results showed that the complaint is caused by external force like dropping lwhich led to a deformation of the adapter upper rim.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORALBPWRRCHGTOOTHBRUSHHANDLE3758IO9SRSM9
Type of Device
TOOTHBRUSH, POWERED - JEQ
Manufacturer (Section D)
BRAUN GMBH WERK MARKTHEIDENFELD
40 baumhofstrasse
marktheidenfeld D-978 28
GM  D-97828
Manufacturer (Section G)
BRAUN GMBH WERK MARKTHEIDENFELD
40 baumhofstrasse
marktheidenfeld D-978 28
GM   D-97828
Manufacturer Contact
mgr. affairs, oral care-see co
mason business center 8700
mason-montgomery rd
mason 45040
MDR Report Key14166483
MDR Text Key299044695
Report Number3000302531-2022-00159
Device Sequence Number1
Product Code JEQ
Combination Product (y/n)Y
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number3758
Device Lot NumberAH04010059
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/21/2022
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/26/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
ORALBPWRPWRORALCARERFLSIOSRS (ORAL-B POWER/POWER O; ORALBPWRPWRORALCARERFLSIOSRS (ORAL-B POWER/POWER O
Patient SexMale
-
-