• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LIMITED ADULT HEATED WALL REUSABLE BREATHING CIRCUIT; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LIMITED ADULT HEATED WALL REUSABLE BREATHING CIRCUIT; BTT Back to Search Results
Model Number 900MR810
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint 900mr810 adult heated wall reusable breathing circuit is currently en-route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Event Description
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that five 900mr810 adult heated wall reusable breathing circuits were damaged during use.There was no patient consequence.
 
Manufacturer Narrative
(b)(4).Method: the complaint (b)(4) adult heated wall reusable breathing circuit was returned to fisher & paykel healthcare in new zealand for investigation, where it was visually inspected.Results: visual inspection revealed that there was helix delamination near the chamber end cuff.There were no signs of tearing or stretching on the tubing.Conclusion: we are unable to determine the cause of the reported event.All 900mr810 adult heated wall reusable breathing circuits are visually inspected and pressure tested prior to being released for distribution.Any circuits that fail are rejected.This suggests the reported damage occurred after the subject breathing circuit was released for distribution.The user instructions that accompany the 900mr810 adult heated wall reusable breathing circuit state the following: - "inspect circuit before re-use, do not use if the circuit shows signs of deterioration, such as: cracks, tears, or damage." - "perform a pressure and leak test on the breathing system, and check for occlusions, before connecting to a patient." - "disconnect tube by handling end connectors only, do not pull or twist tubing, as this may cause damage." - ''do not cover the circuit with materials such as blankets, towels or bed linen.''.
 
Event Description
A healthcare facility in taiwan reported via a fisher & paykel healthcare (f&p) field representative that five 900mr810 adult heated wall reusable breathing circuits were damaged during use.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT HEATED WALL REUSABLE BREATHING CIRCUIT
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618-2216
9494534000
MDR Report Key14183959
MDR Text Key294606754
Report Number9611451-2022-00363
Device Sequence Number1
Product Code BTT
UDI-Device Identifier09420012432247
UDI-Public(01)09420012432247(10)2101384973(11)201112
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K131957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number900MR810
Device Catalogue Number900MR810
Device Lot Number2101384973
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/31/2022
Initial Date FDA Received04/22/2022
Supplement Dates Manufacturer Received05/10/2022
Supplement Dates FDA Received05/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-