Catalog Number CON-HL-90 |
Device Problems
Device Alarm System (1012); Defective Alarm (1014); Break (1069); Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
One device was received and evaluated.Visual inspection confirmed the front cover was broken.Functional testing confirmed the overtemp alarm does not sound and could not be adjusted.The pcb was not functioning properly.The cause of the malfunctioning pcb could not be determined.The most probable root cause of the physical damage to the device is improper handling or storage.
|
|
Event Description
|
It was reported that the front case was damaged along the edges and where it mounts on the bottom.Additionally the high temp and over temp alarms were not alarming.No patient involvement was reported.
|
|
Manufacturer Narrative
|
This mdr was generated under protocol b10010116, as a result of warning letter cms# (b)(4).No causes or potential causes of the customer's reported problem were found during the review of service and repair records.
|
|