• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 595000-001
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Anxiety (2328); Suicidal Ideation (4429)
Event Date 04/12/2022
Event Type  Injury  
Event Description
The customer, a syncardia certified hospital, reported that the patient had suicidal ideation.The hospital coordinated a welfare check with the authorities and the patient came to the hospital for his monthly clinic visit.
 
Manufacturer Narrative
The syncardia 70cc temporary total artificial heart (tah-t) is an implantable pulsatile biventricular replacement device that replaces a patient's native ventricles and valves and pumps blood to both the pulmonary and systemic circulation systems.The syncardia 70cc tah-t is indicated for use as a bridge to transplantation in cardiac transplant-eligible candidates at risk of imminent death from biventricular failure.The syncardia tah-t system is intended for use inside and outside the hospital.The freedom driver is an external driver that provides pneumatic power to the tah-t.The patient was implanted with a tah-t on (b)(6) 2017, however, due to his antibodies, a suitable match is not anticipated.The the patient shared with the hospital and syncardia staff that he is at his wits end because after 5 years of being supported by the syncardia tah-t and freedom driver his marriage has fallen apart, he will be homeless, he has no job or caregiver and is having difficulty getting visas to get a family member into the usa to help him.Patient stated that the noise of the freedom driver is disruptive to his activities of daily living (adls) and he is self-conscious.He cannot go out to eat, socialize, drive or work and everyone stares at him.He wants to go see his family or have his family come see him and it has been a severe burden on him mentally to try to get anything done.Based on the provided information there is no evidence to indicate any device malfunctions or performance issues of the device that would impact the reported event.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key14193698
MDR Text Key289973835
Report Number3003761017-2022-00042
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue Number595000-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/12/2022
Initial Date FDA Received04/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention; Life Threatening;
Patient Age47 YR
Patient SexMale
-
-