Catalog Number 3301-1 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/29/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
|
Event Description
|
It was reported that a closure top stripped during final tightening intra-operatively.There was no reported patient impact.
|
|
Manufacturer Narrative
|
Device evaluation: visual inspection revealed minor damage to the threads and screw drives of the closure tops.Potential cause: root cause was unable to be determined.This event could possibly be attributed to off-axis forces applied during closure top insertion, cross-threading, or using a stripped driver.Dhr review: per dhr reviews, the parts were likely conforming when they left zimvie control.Device use : this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
|
Event Description
|
It was reported that a closure top stripped during final tightening intra-operatively.There was no reported patient impact.
|
|