• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25REC
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2022
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, a connector on the 3 way luer manifold became loose during priming/set up.No patient involvement.Product was changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - updated lot number and added exp date).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #2: 3331 - analysis of production records.Investigation findings: 3259 - improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The returned sample was visually inspected upon receipt.It was confirmed that the lock adaptor was not attached to the manifold.However, the lock adaptor is a press fit and was able to be pushed back onto the manifold.A representative retention sample was inspected for damage with no anomalies noted on the device specifically with the manifold lock adaptor.All manifold units are 100% visually inspected and leak tested during the production process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on april 26, 2022.  upon further investigation of the reported event, the following information is new and/or changed: h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25REC W/RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key14204175
MDR Text Key291313684
Report Number1124841-2022-00095
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450820
UDI-Public(01)00699753450820
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/05/2021
Device Model Number3CX*FX25REC
Device Catalogue NumberN/A
Device Lot NumberZM04
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/07/2022
Initial Date FDA Received04/26/2022
Supplement Dates Manufacturer Received04/28/2022
06/06/2022
Supplement Dates FDA Received05/18/2022
06/10/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-