MEDTRONIC, INC EVEREST INFLATION DEVICE 30ATM; ADAPTOR, STOPCOCK, MANIFOLD, FITTING, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number AC3200
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2022
Event Type  malfunction  
Event Description
An attempt was made to use one everest inflation device during a procedure to treat a non-tortuous and non-calcified lesion located in the pulmonary artery.The device was inspected with no issues noted.The device was prepped per ifu with no issues noted.The indeflator was connected to three balloon devices which were in the patient's body.It was reported that balloon rupture occurred with each device.Pressure was applied from the range of nominal burst pressure (nbp) to rated burst pressure (rbp), however balloon ruptures occurred.It was stated that the actual pressure applied may have been higher than the pressure value that was displayed on the pressure gauge.The patient is alive with no injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: negative prep was performed on the device prior to use with no issues noted.The lesion was pre-dilated.A stopcock was connected between the everest device and the balloons.No resistance was noted while advancing the balloons to the lesion.The balloons were not moved or repositioned in the lesion while inflated.Two of the balloons were non-medtronic balloons and one was an unknown balloon.Different deflators, which were not everest devices, were also used.The everest device was associated with all three bursts.It is not known which balloon ruptured at what pressure, but the balloon was ruptured at 8atm for nominal and 14atm for rated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis summary: one everest inflation device returned for analysis.No damage was noted to the syringe body, tube joints or manometer housing.As received the gauge needle was at approx.0 atm.Using an inhouse stopcock, the device was pressurized with water.The gauge needle moved steadily to 30 atm and the device held pressure for a minimum of three minutes (with max 10% loss), meeting specification.Upon release of pressure the gauge needle dropped to approx.0 atm.When vacuum was applied, the gauge needle moved into the red "vac" reading.An in-house euphora device was inflated, using the returned device, to a nominal of 12atm and burst rate of 14atm with no issues noted.No other damage evident to the remainder of the device.Standalone medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: annex d code medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVEREST INFLATION DEVICE 30ATM
Type of Device
ADAPTOR, STOPCOCK, MANIFOLD, FITTING, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
Manufacturer (Section G)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14217132
MDR Text Key295246349
Report Number1220452-2022-00033
Device Sequence Number1
Product Code DTL
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K153038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/29/2023
Device Catalogue NumberAC3200
Device Lot Number60328295
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-