Model Number 2426-0500
Device Problem Device Displays Incorrect Message (2591)
Patient Problem Insufficient Information (4580)
Event Date 04/05/2022
Event Type  malfunction  
Event Description
The rn was attempting to infuse medication, and the pump continually stated "air in line" after multiple attempts to trouble shoot with multiple rns.Once the tubing was changed, the infusion ran smoothly.The tubing was from pharmacy, and they were unsure of the exact tubing packaging it came from.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PUMP INFUSION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION 303, INC.
10020 pacific mesa blvd
san diego CA 92121
MDR Report Key14218343
MDR Text Key290171022
Report Number14218343
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number2426-0500
Device Catalogue Number2426-0500
Device Lot Number21103419
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/11/2022
Event Location Outpatient Treatment Facility
Date Report to Manufacturer04/27/2022
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/27/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age10585 DA
Patient SexMale
-
-