• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST E PLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST E PLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 875205
Device Problem Overheating of Device (1437)
Patient Problems Burn(s) (1757); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional serial numbers included in this report: (b)(4).28 of 29 devices were returned for evaluation.1 device will not be returned for evaluation.Evaluation of one device found it to be within specification.Evaluation of one device found excessive wear due to a lack of proper maintenance and lubrication.The device did not heat up during evaluation.The device had debris build-up.The device was repaired and returned to the customer.Evaluation of 24 devices found excessive wear due to a lack of proper maintenance.The devices did not heat up during evaluation.The devices had debris build-up.The devices were repaired and returned to the customers.Evaluation of 2 devices found excessive wear due to a lack of proper maintenance.The devices did heat up during evaluation.The devices had debris build-up.The devices were repaired and returned to the customers.
 
Event Description
This report summarizes 29 malfunction events where a midwest e plus handpiece overheated.In 4 events, the event outcome is unknown despite reasonable follow-up.In 3 events, patients experienced minor burns that did not require intervention.No injury resulted in 22 of the events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST E PLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key14219404
MDR Text Key290453219
Report Number9614977-2022-00013
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K972436
Number of Events Reported29
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number875205
Patient Sequence Number1
-
-