Model Number 4FC12
Device Problem Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2022
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the 4fc12 sheath with lot 0010925125 was returned and analyzed.Visual inspection of the handle area was performed.The inspection identified that the stopcock was detached from the side port tube; the stopcock was not returned.Insertion of dilator into the sheath was performed.It was not possible to lock the dilator luer into the sheath.Further inspection under a microscope identified dilator luer damage.In conclusion, the sheath failed the returned product inspection due to a detached stopcock from the sheath.If information is provided in the future, a supplemental report will be issued.
 
Event Description
After a completed case, the sheath subsequently tested out of specification per the manufacturers investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14220027
MDR Text Key290173178
Report Number3002648230-2022-00188
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/11/2023
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0010925125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Initial Date Manufacturer Received 03/09/2022
Initial Date FDA Received04/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-