Model Number HL-90
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the device keeps alarming.No patient involvement was reported.
 
Event Description
Additional information received, indicated that this occurred when inserting the tubing into the machine.
 
Manufacturer Narrative
Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found broken light-emitting diodes (led), cracked front cover, and dirty heater assembly.Wear and tear damaged line cord.Functional testing found the over temp alarm keeps going off confirming the customer complaint - alarm was out of calibration and the potentiometer (pot) used to set it was unresponsive to adjustments.Replaced the printed circuit board (pcb), heater assembly, front cover, and line cord.The device passed all functional testing.The root cause of the reported issue was found to be component failure.D5 and e4 are unknown, no information provided to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
suite 300
minneapolis, MN 55442
MDR Report Key14221239
MDR Text Key290184950
Report Number3012307300-2022-06991
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberHL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-