Model Number AZURION 5 M12
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/13/2022
Event Type  malfunction  
Event Description
It has been reported to philips that during a procedure the l-arm rotation cover became detached.No harm has been reported to philips.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Additional narrative: philips has investigated this complaint.According to the information collected, during a diagnostic procedure, the l-arm cover detached after a collision with the operation light.The cover was retained by the safety chain and the procedure was completed as planned.A philips service engineer inspected the system onsite and identified that the cover could not be reseated due to a crack.The engineer replaced and installed the l-arm cover and returned the system to use in good working order.Corrected data: codes were updated as per the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key14235378
MDR Text Key290274601
Report Number3003768277-2022-00183
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00884838099227
UDI-Public00884838099227
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K200917
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAZURION 5 M12
Device Catalogue Number722227
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-