Related manufacturer reference number: 1627487-2022-02454, 1627487-2022-02455.It was reported that the patient experienced loss of therapy from one side.During following up with the patient on (b)(6) 2022, diagnostics revealed that there were low impedances.Reprogramming was unable to restore therapy.In turn, surgical intervention took place on (b)(6) 2022.During the procedure it was noted that the extensions were heavily intertwined and one of the extensions was broken.Reportedly, the patient rotated the ipg subcutaneously.As such, the extensions were explanted and replaced and the ipg was repositioned to its correct orientation addressing the issue.Reportedly, therapy was restored post operatively.
|