Model Number N/A
Device Problems Activation Failure (3270); Activation Problem (4042)
Patient Problem Needle Stick/Puncture (2462)
Event Type  malfunction  
Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Health professional: rn.The root cause of the complaint could not be identified to be related to our production or process.The actual sample was not available for our evaluation.Retention samples were visually confirmed free from defects that will affect activation of safety sheath and passed evaluation for sheath activation and deactivation.Also, no irregularity was encountered during the simulation of manual sheath activation that may lead to the complaint.We have a series of visual in-process inspections to detect an abnormality on the sheath that may lead to a problem during sheath activation.The molding condition of the components critical to the safety activation of the product is routinely checked to assure that no defects will be encountered that will lead to sheath activation problems and needle sticks.Similarly, the assembly status of the safety needle such as sheath collar fitting and damaged parts that will affect product function is being confirmed.Lot history file revealed no related nonconformity or irregularity that will lead to the complaint.Prior to shipment, qc conducts outgoing visual, sensory, and functional inspections to assure lots are of good quality.Therefore, we advise to follow the instructions for use (ifu) for the proper usage of the sg2 needle indicated on the unit box in which warnings to prevent needle stick, cautions, and precautions are also included.Terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of terumo ((b)(4)) corporation (manufacturer) registration no.(b)(4).
 
Event Description
The user facility reported that the safety engagement malfunction led to a needle stick.The staff member thought the safety was engaged; however, it was not.Additional information was received on 31 mar 2022: the reported product was used for a patient injection.When the staff member attempted to engage the safety, they thought they heard the click, as intended for successful containment of needle into sheath.The staff member incurred a needle stick because the device was not engaged, and the exposed needle stuck them.The needle stick injury did not require medical intervention.A simple first aid wash/band-aid was applied.The patient in the procedure room was not impacted in any way.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON PREVENT ® HT SAFETY HYPODERMIC NEEDLE
Type of Device
SYRINGE, ANTISTICK
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan,
RP 
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna,,
RP  
Manufacturer Contact
stephanie handy
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key14241163
MDR Text Key292519592
Report Number3003902955-2022-00009
Device Sequence Number1
Product Code MEG
UDI-Device Identifier30612479178855
UDI-Public30612479178855
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number102-N251S
Device Lot Number210730B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial