Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed scratches on the top and bottom covers, a dent on the bottom cover, as well as a severed electrode.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical test performed.This device was explanted for medical reasons.However, this device had moisture that exceeded the residual gas analysis (rga) test limit.Based on an assessment of the rga data, it is determined that this device was non-hermetic.Corrective actions were implemented.This version of the 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|