• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST PHOENIX; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST PHOENIX; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 791430
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
This report summarizes 2 malfunction events where a midwest phoenix handpiece would not hold burs.No injury resulted in any of the events.
 
Manufacturer Narrative
2 of 2 devices were returned for evaluation.Evaluation of 1 device found normal chuck wear and the turbine needed to be replaced.The device was repaired and returned to the customer.Evaluation of 1 device found chuck wear, lack of maintenance, and lack of lubrication.The device was cleaned of debris and the turbine was replaced and the device was properly lubricated.The device was repaired and returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST PHOENIX
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key14251668
MDR Text Key290452465
Report Number9614977-2022-00022
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K131319
Number of Events Reported2
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number791430
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/29/2022
Patient Sequence Number1
-
-