• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Communication or Transmission Problem (2896)
Patient Problem Insufficient Information (4580)
Event Date 02/13/2021
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of electrode belt has been completed.The reported problem (unable to communicate with monitor) has been confirmed.Upon investigation the trunk cable connector was physically damaged and the electrode belt was unable to securely connect to a monitor.The root cause for the damaged connector was excessive force.No adverse event resulted from the damaged belt.
 
Event Description
A us distributor reported that a patient's electrode belt was returned for investigation into a non-reportable patient death and was unable to communicate with a monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
landyn blinn
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key14254642
MDR Text Key295413801
Report Number3008642652-2022-04159
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005012
UDI-Public00855778005012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0987
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2022
Date Manufacturer Received04/04/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-