The investigation of the returned coil system found the hypotube bent, the body coil kinked, and the implant not attached to the pusher.No indication using a detachment controller was found on the pusher's heater coil.However, further inspection found that the implant was stuck inside of the dispenser hoop.The investigation found the implant's monofilament with a tensile break shape at the tip which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
|