Model Number K-LESS T-ROPE W/DRV, SYN REPR, TI
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2022
Event Type  malfunction  
Event Description
On (b)(6) 2022, it was reported by a sales representative via email that a total of four ar-8925t syndesmosis tightrope broke at the round lateral button during the final tightening.Surgeon removed all failed hardware and had to make a medial incision in order to retrieve the buttons.This caused a significant delay in the procured.Case was completed using a syndesmosis screw.This was discovered during an ankle fracture procedure on (b)(6) 2022 additional information received on 4/8/2022: all four ar-8925t syndesmosis tightrope broke and all were removed from inside the patient.No additional anesthesia had to be administered to patient due to delay in procedure.
 
Manufacturer Narrative
The complaint is confirmed based on the customer-provided photo, which displays the three broken suture constructs.However, the cause remains undetermined without receipt of the devices for physical evaluation.No change in harm was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
K-LESS T-ROPE W/DRV, SYN REPR, TI
Type of Device
WASHER, BOLT NUT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14274541
MDR Text Key299569626
Report Number1220246-2022-04848
Device Sequence Number1
Product Code HTN
UDI-Device Identifier00888867281608
UDI-Public00888867281608
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043248
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberK-LESS T-ROPE W/DRV, SYN REPR, TI
Device Catalogue NumberAR-8925T
Device Lot Number14570053
Was Device Available for Evaluation? No
Date Manufacturer Received04/07/2022
Date Device Manufactured11/19/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-