BD MEDICAL - DIABETES CARE BD TUBERCULIN SYRINGE WITH NEEDLE SAFETYGLIDE¿; ANTISTICK SYRINGE Back to Search Results
Model Number 305945
Device Problem Break (1069)
Patient Problems Needle Stick/Puncture (2462); Foreign Body In Patient (2687)
Event Date 04/06/2022
Event Type  Injury  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported that bd tuberculin syringe with needle safetyglide¿ broke off in the patient and the health care provider was stuck with the dirty needle.It was reported needle broke off.Event date: (b)(6) 2022.Awareness date: 7-apr-2022.Complaint: the syringe broke off in the patient, the needle stuck the rn.
 
Manufacturer Narrative
H6: investigation summary: customer returned a photo of a 1ml safetyglide syringe.Customer states that the needle broke off.The photo was examined and exhibited a broken plunger rod.A review of the device history record was completed for batch #1257281.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Embecta was able to duplicate or confirm the customer¿s indicated failure.Dhr and l2l dispatches were reviewed, nothing pertaining to this defect was found.No root cause can be found.
 
Event Description
It was reported that bd tuberculin syringe with needle safetyglide¿ broke off in the patient and the health care provider was stuck with the dirty needle." it was reported needle broke off." verbatim: event date: 6-apr-2022.Awareness date: 7-apr-2022.Complaint: the syringe broke off in the patient, the needle stuck the rn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD TUBERCULIN SYRINGE WITH NEEDLE SAFETYGLIDE¿
Type of Device
ANTISTICK SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14274927
MDR Text Key293100018
Report Number1920898-2022-00307
Device Sequence Number1
Product Code MEG
UDI-Device Identifier30382903059455
UDI-Public30382903059455
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K992734
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number305945
Device Catalogue Number305945
Device Lot Number1257281
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-