DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/08/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the touchscreen for the cardiosave intra-aortic balloon pump (iabp) was not functional.It is unknown under which circumstances this event occurred; however there was no patient involvement, and no adverse event was reported.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and observed that the touchscreen was out of calibration.The fse completed a touchscreen calibration then verified that the touchscreen was working.Subsequently, the fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that the touchscreen for the cardiosave intra-aortic balloon pump (iabp) was not functional.This event occurred during a pre-check, there was no patient involvement.
|
|