• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Communication or Transmission Problem (2896)
Patient Problem Insufficient Information (4580)
Event Date 04/22/2022
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor has been completed.The reported problem (does not communicate with belt) has been confirmed.Upon investigation the monitor was unable to communicate with an electrode belt.The monitor's electrode belt connector was broken free from the monitor enclosure, damaging wires within the connector.The root cause for the damaged connector was excessive force.No adverse event resulted from the damaged monitor.
 
Event Description
A us distributor returned a monitor and reported the monitor was unable to communicate with an electrode belt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
landyn blinn
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key14286810
MDR Text Key296422469
Report Number3008642652-2022-04641
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0988
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2022
Date Manufacturer Received04/29/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-