The braun/welch allyn thermoscan pro 6000 ear thermometer is indicated for the intermittent measurement of human body temperature for patients having ages ranging from normal weight (full term) newborn to geriatric adults in a professional use environment.The probe cover is used as a sanitary barrier between the infrared thermometer and the ear canal.To help ensure accurate temperature measurements, the sensor itself is warmed to a temperature close to that of the human body.When the braun/welch allyn thermoscan is placed in the ear, it continuously monitors the infrared energy until a temperature equilibrium has been reached and an accurate measurement can be taken.Hillrom/welch allyn¿s investigation of pro 6000 ¿hot tip¿ complaints confirmed that aggressive cleaning and decontamination practices can cause liquid ingress.Consequently, liquid ingress can adversely affect the temperature sensor causing the pro 6000 device to behave inconsistently and to overheat the speculum tip.It is believed that replication of the malfunction with the returned devices from the customer at the manufacturing site has not been possible as any fluid that may have ingressed had likely evaporated, therefore not showing the ¿hot tip¿ malfunction.Based on hillrom¿s ability to replicate the malfunction of a ¿hot tip¿ on new devices that can potentially go above the built in risk mitigation of a safety cut off that could potentially cause a more serious injury we have deemed this complaint to be reportable.
|