Model Number HL 20
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2022
Event Type  malfunction  
Manufacturer Narrative
It was reported that the error message "belt slip" was displayed on two hl 20 pumps.No harm to any person was reported.A getinge field service technician was onsite and investigated the unit in question.The technician calibrated the tacho board for both pumps, after that the pumps did not display the error message anymore.The device was checked for two hours without any issue.Hl20 device is back in use working to factory specification.No parts were replaced.Thus the reported failure could be confirmed.However the failure mode "belt slip" can be linked to the following most possible root causes according to the hl 20 risk management file ((b)(4) version 11): failure of pump control board, defective/ dirty tacho, relay or pump belt.The device was manufactured in 2014-03-16.The review of the non-conformities during the period of 2014-03-16 to 2022-04-28 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that 2 hl20 pumps were displaying the error message: "beltslip".No patient involved.Reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key14287192
MDR Text Key290781435
Report Number8010762-2022-00148
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number70102.7652
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2014
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-