There is a plan to have the erbe equipment checked/evaluated at erbe usa.If any issues are found, a follow up report will be provided.No anomalies were found in device history record (dhr) of the involved device.Most likely, there were many factors involved in the event.At this time, no determination could be made as to the cause of the incident.
|
It was reported that a patient incident occurred with the electrosurgical unit (esu/generator) during a colonoscopy to perform a polypectomy.The esu was used with a 13 mm boston snare.A description of the other accessories used in the procedure was not provided.The generator settings were as well as forced coag, effect 2, 20 watts and endocut q, settings 2, 1, and 6.Specifically, it was documented by the physician, dr.(b)(6), "during colonoscopy (i was doing the procedure), large polyp found on the hepatic flexure.Injected to lift polyp without issues.Once hot snare was used, the 'coag' pedal functioned appropriately but the 'cut' pedal did not.The tech was able to snare the polyp but it was done without the cut function.After the polyp was removed, a perforation was noted at the polypectomy site.This was successfully closed using clips, and the patient was admitted to the hospital for observation and antibiotics.This perforation likely occurred due to a device malfunction." no further information was provided in regards to the patient's condition.As a result of the incident, the medical center's clinical engineering department checked the unit.The esu passed their safety and performance test.
|