Model Number N/A
Device Problems Failure to Cut (2587); Patient Device Interaction Problem (4001)
Patient Problem Abrasion (1689)
Event Date 03/22/2022
Event Type  Injury  
Event Description
It was reported that the device does not remove skin evenly; it skims or skips in steps as if jumping up and down as it is moved across the skin.The event timing was during surgery.There was harm to the patient and an additional skin graft was needed.No additional patient consequences occurred as a result of this malfunction.
 
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
 
Event Description
There is no additional event details available.
 
Manufacturer Narrative
Review of the most recent repair record determined the unit was confirmed to not cut properly.The motor speed was low and the unit was out of calibration.The motor, handpiece switch, bearing pack, oring, seal, reciprocating arm, and ss screw were replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME, LINE-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key14288606
MDR Text Key290802630
Report Number0001526350-2022-00430
Device Sequence Number1
Product Code GFD
UDI-Device Identifier0088902437599
UDI-Public(01)0088902437599(11)180918(10)64155055
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number64155055
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received05/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-