Model Number 27093
Device Problem Operating System Becomes Nonfunctional (2996)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned to an authorized resmed third party service center and an evaluation confirmed the complaint.The top case was replaced to address this issue.The device was serviced and fully tested before it was returned to the customer.Resmed reference #: (b)(4).
 
Event Description
It was reported to resmed that an astral device had an unresponsive touchscreen.There was no patient harm or serious injury reported as a result of this incident.
 
Event Description
It was reported to resmed that an astral device had an unresponsive touchscreen.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to a manufacturing issue.Resmed¿s risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - FUKUDA
Type of Device
VENTILATOR, CONTINUOUS (FACILITY/HOME)
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key14291904
MDR Text Key296687821
Report Number3004604967-2022-00314
Device Sequence Number1
Product Code NOU
UDI-Device Identifier00619498270934
UDI-Public(01)00619498270934(10)1516089
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K172875
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation 505
Type of Report Initial,Followup
Report Date 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27093
Device Catalogue Number27093
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2022
Distributor Facility Aware Date02/13/2023
Date Manufacturer Received02/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A