Model Number 21-7302-24
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2020
Event Type  malfunction  
Manufacturer Narrative
A dhr review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Visual inspection was performed on 4 pictures and 8 samples received in new conditions inside their closed original packaging.No particles were found in the pictures received.2 of the returned samples no contamination was observed.6 remaining samples the threshold of contamination was not observed.No root cause could be determined since the complaint was not confirmed because samples meet the specification.No corrective actions were taken since the complaint was not confirmed.
 
Event Description
It was reported foreign material speck between bladder and wall of cassette.No patient involvement was reported.
 
Manufacturer Narrative
While performing a review of submitted mdr report it was discovered that this file was inadvertently assessed as reportable.No patient death, serious injury, and no evidence of a reportable product malfunction.The hazardous situation for the given complaint device would not likely cause or contribute to a death or serious injury if the malfunction were to recur.File (b)(4) is no longer considered reportable, please disregard any mdr reports associated with it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD CASSETTE RESERVOIRS - FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14311820
MDR Text Key290960407
Report Number3012307300-2022-07874
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027239
UDI-Public10610586027239
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number21-7302-24
Device Catalogue Number21-7302-24
Device Lot Number3941138
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-0148-2022
Patient Sequence Number1
-
-