MEDOS INTERNATIONAL SÃ RL CH UNKNOWN LOCKING/SET SCREWS; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR
|
Back to Search Results |
|
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is for an unknown locking/set screws /unknown lot.Part and lot number are unknown.Without the specific part number; the udi number and 510-k number is unknown.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j employee.Device evaluated by mfr and manufacture date: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, during a plif (the lumbosacral transitional vertebrae) for the treatment lss and sacrum spine facture, the surgeon tried to tighten the rod connector with the setscrew, the setscrew was stripped.Another device was used to complete the procedure.The procedure was completed less than 30-minute surgical delay.The procedure was successfully completed.This report involves one (1) unknown locking/set screws.This is report 2 of 2 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D10: date of concomitant therapy is (b)(6) 2022.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was further reported that patient was stable.
|
|