• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD ADMINISTRATION SETS WITH FLOW STOP FREE-FLOW; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD ADMINISTRATION SETS WITH FLOW STOP FREE-FLOW; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7322-24
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2021
Event Type  malfunction  
Event Description
It was reported that lines would not prime.No patient injury was reported.
 
Manufacturer Narrative
Other, other text: visual inspection and functional testing was performed on one sample returned.No obstructions nor other workmanship defects were detected in none of the joins of the product.No discrepancies were detected the product was prime without difficult, liquid flow through whole devices; thus the failure mode reported is not confirmed.No corrective actions are required since the complaint was not confirmed.A dhr review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
 
Manufacturer Narrative
While performing a review of submitted mdr report it was discovered that this file was inadvertently assessed as reportable.No patient death, serious injury, and no evidence of a reportable product malfunction.The hazardous situation for the given complaint device would not likely cause or contribute to a death or serious injury if the malfunction were to recur.Report 3012307300-2022-08049 is no longer considered reportable, please disregard any mdr reports associated with it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SETS WITH FLOW STOP FREE-FLOW
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14321231
MDR Text Key291315338
Report Number3012307300-2022-08049
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586029646
UDI-Public10610586029646
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse Practitioner
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7322-24
Device Catalogue Number21-7322-24
Device Lot Number4010271
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/22/2022
Initial Date FDA Received05/07/2022
Supplement Dates Manufacturer Received06/06/2022
Supplement Dates FDA Received06/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-0148-2022
Patient Sequence Number1
-
-