MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CARDIOHELP
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2022
Event Type  malfunction  
Event Description
A defective arteriel bubble sensor (flow/bubble sensor) was reported.No harm to any person has been reported.New information received on 2022-05-05, that the cardiohelp was exchanged during treatment.Complaintnumber: (b)(4).
 
Manufacturer Narrative
A follow up will be submitted when additional information become available.A getinge technician will investigate the cardiohelp in question.
 
Manufacturer Narrative
A defective arterial bubble sensor (flow/bubble sensor) was reported.Customer reported, that the bubble sensor was not reading any values.The time of occurence was not provided.A getinge service technician (fst) was sent for investigation and repair on 2022-04-26/27.The flow/bubble sensor was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.It was confirmed by fst, that most probably customer used non getinge parts.The review of the non-conformities has been performed on 2022-05-16 for the period of 2012-10-01 to 2022-04-26.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely to have contributed to the reported failure.Based on the results the reported failure "defective arterial bubble sensor (flow/bubble sensor)" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key14335137
MDR Text Key291244178
Report Number8010762-2022-00157
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP
Device Catalogue Number701048012
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received07/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-