• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. CARESITE®; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. CARESITE®; SET, ADMINISTRATION, INTRA Back to Search Results
Model Number 354205
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2022
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As reported by the user facility: detailed inquiry description: it was reported that the iv administration set and care site have oil on the connector.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number 400554118.One (1) unused sample in opened packaging was returned for evaluation.A visual evaluation of the returned sample did not show any signs of grease/oil/dirt or any foreign substance on the valves, connectors, or any other component on the set.Based on the evaluation results, the reported defect was not able to be confirmed.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESITE®
Type of Device
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer (Section G)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
santo domingo
DR  
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key14348207
MDR Text Key291371240
Report Number2523676-2022-00191
Device Sequence Number1
Product Code FPA
UDI-Device Identifier04046964182099
UDI-Public(01)04046964182099
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number354205
Device Catalogue Number470124
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/28/2022
Initial Date FDA Received05/10/2022
Supplement Dates Manufacturer Received04/28/2022
Supplement Dates FDA Received06/16/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-