• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG Back to Search Results
Model Number 6662
Device Problem Migration (4003)
Patient Problem Implant Pain (4561)
Event Date 04/20/2022
Event Type  Injury  
Manufacturer Narrative
Event date is estimated.Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Event Description
It was reported that the patient¿s experienced discomfort due to ipg moving down.The ipg was repositioned on (b)(6) 2022 due to bow stringing of the extension in the neck.During the surgery it was noted that the plastic suture was broken due to which the sutures did not hold.As such, the generator was re-secured addressing the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR
Type of Device
DBS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key14349637
MDR Text Key291342924
Report Number1627487-2022-02688
Device Sequence Number1
Product Code MHY
UDI-Device Identifier05415067020260
UDI-Public05415067020260
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model Number6662
Device Catalogue Number6662
Device Lot Number8068955
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DBS BURRHOLE CAP; DBS BURRHOLE CAP; DBS EXTENSION (X2); DBS LEAD; DBS LEAD
Patient Outcome(s) Other;
Patient SexFemale
Patient Weight50 KG
-
-