• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KINETIC CONCEPTS, INC. ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM; OMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KINETIC CONCEPTS, INC. ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM; OMP Back to Search Results
Model Number WNDARM
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Sepsis (2067)
Event Date 04/15/2022
Event Type  Injury  
Event Description
On 19-apr-2022, the following information was reported to kci by the home health case manager: the patient was admitted to the hospital on (b)(6) 2022 allegedly due to sepsis.No additional information was provided.On 31-mar-2022, the device was tested per quality control procedure by the kci service center, and the unit passed the quality control checks and met specifications.On (b)(6) 2022, the device was placed with the patient.Multiple unsuccessful attempts have been made to retrieve the device; therefore, a device evaluation could not be performed.
 
Manufacturer Narrative
Based on the information provided, it cannot be determined that the alleged sepsis is related to the activ.A.C.¿ ion progress¿ remote therapy monitoring system.Multiple unsuccessful attempts have been made to obtain additional clinical information.It is unknown if and what medical or surgical intervention was required.The device passed quality control checks before patient placement, and a device evaluation after placement could not be performed.Device labeling, available in print and online, states: wound infection call your doctor or nurse right away if you think your wound is infected or if the following symptoms develop or worsen: you have a fever.Your wound is sore, red or swollen.Your skin itches or you have a rash or redness around the wound.The area around the wound feels very warm.You have pus or a bad smell coming from the wound.Infected wounds: infected wounds should be monitored closely and may require more frequent dressing changes than noninfected wounds, dependent upon factors such as wound conditions, treatment goals.Refer to dressing application instructions (found in v.A.C.® dressing cartons) for details regarding dressing change frequency.As with any wound treatment, clinicians and patients / caregivers should frequently monitor the patient's wound, periwound tissue and exudate for signs of infection, worsening infection, or other complications.Some signs of infection are fever, tenderness, redness, swelling, itching, rash, increased warmth in the wound or periwound area, purulent discharge or strong odor.Infection can be serious, and can lead to complications such as pain, discomfort, fever, gangrene, toxic shock, septic shock and/or fatal injury.Some signs or complications of systemic infection are nausea, vomiting, diarrhea, headache, dizziness, fainting, sore throat with swelling of the mucus membranes, disorientation, high fever, refractory and/or orthostatic hypotension or erythroderma (a sunburn-like rash).If there are any signs of the onset of systemic infection or advancing infection at the wound site, contact the treating physician immediately to determine if v.A.C.® therapy should be discontinued.
 
Manufacturer Narrative
Based on the additional information provided regarding the device, kci's assessment remains the same; it cannot be determined that the alleged sepsis is related to activ.A.C.¿ ion progress¿ remote therapy monitoring system.The device passed quality control checks before and after patient placement.
 
Event Description
On 20-jun-2022, a device evaluation was completed by kci quality engineering.On 20-jun-2022, the device was tested per quality control procedure by kci service center, and the unit passed the quality control checks and met specifications.Inspection and testing of the device did not reveal any evidence of an operational malfunction with the unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM
Type of Device
OMP
Manufacturer (Section D)
KINETIC CONCEPTS, INC.
san antonio TX 78249
Manufacturer Contact
steven jackson
6203 farinon drive
san antonio, TX 78249
2102556438
MDR Report Key14359654
MDR Text Key291409953
Report Number3009897021-2022-00091
Device Sequence Number1
Product Code OMP
UDI-Device Identifier00849554005600
UDI-Public0100849554005600
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberWNDARM
Device Catalogue Number420095
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age46 YR
Patient SexFemale
-
-