The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.Lot and serial number of the device not provided by the complainant; therefore, the udi, expiration and manufacturing dates are not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.
|
It was reported that during a myosure procedure on (b)(6) 2022, a uterine perforation that included bowel perforation was observed.The patient was (b)(6), cavity full with polyps, after the procedure the patient did not feel well and got a fever.The results from pathology came back as bowel tissue.No other information is available at the moment.
|