• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC CONFIDENCE SPINAL CMT SYS, 11C; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC CONFIDENCE SPINAL CMT SYS, 11C; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Model Number 283910000
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Event Description
It was reported from taiwan that the confidence spinal cmt device seal at the column is not complete and it would leak.It is unknown if there was patient involvement.This complaint involves one (1) device confidence spinal cmt sys, 11c.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j sales representative.Device evaluated by mfr and manufacture date: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9 e1 h3, h6: a manufacturing record evaluation was performed for the finished device product code: 283910000 lot number: 329342 it was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 22.11.2021 cement number: product code : 183901001 / batch number : 9857920 the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there are no defects on the surface of the confidence spinal cmt sys, 11c.A dimensional inspection was unable to be performed due to device design.A functional test was performed to determine if the device was leaking.After turning the pump clockwise, water started leaking from the bottom of the pump body.The complaint condition was replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the confidence spinal cmt sys, 11c would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIDENCE SPINAL CMT SYS, 11C
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14363312
MDR Text Key295621996
Report Number1526439-2022-00692
Device Sequence Number1
Product Code NDN
UDI-Device Identifier10705034209623
UDI-Public10705034209623
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K060300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283910000
Device Catalogue Number283910000
Device Lot Number329342
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-