On the 12th of april 2022, the customer from (b)(6) reported that some slides were affected.It was suggested that the customer re-slices and re-stains the affected slides and stops using the instrument.The fse inspected the instrument and found that the liquid dripped out from the probe.The aspiration and dispense check valve was replaced, the new spare part was tested in diagnostics per service manual.The instrument has been repaired, it is fully operational and is available to the user.No harm or misdiagnosis was indicated.
|