• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CELESTICA DE MONTERREY S.A. DE C.V. ENDO CLIP; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CELESTICA DE MONTERREY S.A. DE C.V. ENDO CLIP; CLIP, IMPLANTABLE Back to Search Results
Model Number 176620
Device Problems Difficult to Open or Close (2921); Fail-Safe Problem (2936); Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2022
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic procedure, in ligation of blood vessels, after four to five clips were fired, the jaws was not able to close.The clips were not loading properly into the jaws.To resolve the issue, the surgeon used another product to complete the case.There was no patient injury.
 
Manufacturer Narrative
Additional information: b5, g3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic procedure, in ligation of blood vessels, after four to five clips were fired, the jaws was not able to close.The clips were not loading properly into the jaws.The white safety lock deployed as well.To resolve the issue, the surgeon used another product to complete the case.There was no patient injury.
 
Manufacturer Narrative
Additional information: d9, g1, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted buckled pusher bar on device due to excessive force applied to the jaws by the end user.It was reported that the clips did not load properly into the jaws as expected, the safety interlock deployed prior to all staples or clips firing, and the jaws of the reload did not close at all.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur under the following conditions: if the load indicator printed on the distal end of the instrument shaft has not cleared the distal end of the cannula and the instrument is fired.If a heavy load is applied to the side of the jaw which partially closes it during the clip loading process.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: attempts to load clips into the device jaw without full clearance of the trocar sleeve, and full visibility to the word ¿load¿ can result in device failure, malfunction, and/or malformed clips.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
CELESTICA DE MONTERREY S.A. DE C.V.
calle octava #102
parque industrial monterrey
apodaca nuevo leon
MX 
Manufacturer (Section G)
CELESTICA DE MONTERREY S.A. DE C.V.
calle octava #102
parque industrial monterrey
apodaca nuevo leon
MX  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key14377726
MDR Text Key292067965
Report Number3003139373-2022-00126
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057876
UDI-Public10884521057876
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K143644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number176620
Device Catalogue Number176620
Device Lot NumberT9L091X
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/27/2022
Initial Date FDA Received05/12/2022
Supplement Dates Manufacturer Received05/27/2022
12/20/2022
Supplement Dates FDA Received06/13/2022
01/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-