• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFESCAN EUROPE GMBH OT VERIO TEST STRIP; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE GMBH OT VERIO TEST STRIP; GLUCOSE MONITORING SYS/KIT Back to Search Results
Lot Number 4802438
Device Problem Material Twisted/Bent (2981)
Patient Problems Hypoglycemia (1912); Shaking/Tremors (2515)
Event Date 04/29/2022
Event Type  Injury  
Event Description
On (b)(6) 2022, the lay user/patient contacted lifescan (lfs) (b)(6), alleging that their onetouch verio test strips were bent and only partially filled after a sample was applied.The complaint was classified based on the customer care agent (cca) documentation.The patient reported that alleged issues began (b)(6) 2022.The patient stated that they are sometimes unable to obtain a blood glucose reading due to the alleged issues.The patient claimed they developed symptom of ¿shaking¿ between 4:00 pm and 5:00 pm before dinner almost every day since (b)(6) 2022; symptom they associated with a low blood glucose.The patient claimed they attempted to test their blood glucose in response to the symptom however quickly self-treated with food and drink.The patient informed the cca that they manage their diabetes with a fixed dose of insulin (3 doses of rapid-acting insulin; 1 dose of lantus) and skipped insulin as a result of the alleged issue.At the time of troubleshooting, the cca noted that the subject meter was not being used for the first time and the correct test strips were being used for testing.The patient was unwilling to walk through resolving the issue.Replacement product was sent to the patient.This complaint is being reported because the patient reportedly developed symptoms suggestive of a serious injury adverse event after the alleged issues began.The subject test strips could not be ruled out as a contributor to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO TEST STRIP
Type of Device
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
gubelstrasse 34
zug 6300
SZ  6300
MDR Report Key14382205
MDR Text Key294263269
Report Number2939301-2022-03037
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Lot Number4802438
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date05/06/2022
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/12/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-