• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1601-04
Device Problems Expulsion (2933); Material Protrusion/Extrusion (2979)
Patient Problems Erosion (1750); Unspecified Infection (1930); Skin Erosion (2075)
Event Type  Injury  
Event Description
The recipient is reportedly experiencing an infection followed by extrusion.Th recipient is presenting with skin erosion.The recipient's device was explanted.The recipient was not reimplanted.
 
Manufacturer Narrative
The recipient has reportedly healed following explant surgery and the issues leading to explantation have resolved.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed tool damage in the silicone on the top and bottom covers, as well as a severed electrode.These are believed to have occurred during revision surgery.The photographic imaging inspection revealed broken electrode wires in the fantail region.This is believed to have occurred during revision surgery.System lock was verified.The condition of the electrode prevented and electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
jennifer rhudy
28515 westinghouse place
valencia, CA 91355
MDR Report Key14402047
MDR Text Key291702633
Report Number3006556115-2022-00614
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016877995
UDI-Public(01)07630016877995(11)200922(17)230930
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/30/2023
Device Model NumberCI-1601-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/13/2022
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received06/10/2022
07/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-