Catalog Number 09211101190
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2011
Event Type  malfunction  
Event Description
In light of the covid-19 pandemic and the subsequent emergency use authorizations (euas) for sars-cov-2 diagnostic tests, the agency has requested heightened reporting beyond the reasonably suggests requirements of 803 to include allegations of false positive or false negative results independent of harm or malfunction or off-label use.Pursuant to the agency¿s instruction, we hereby submit this mdr.A customer from the us alleged a discrepant result for a single patient while using the cobas® sars-cov-2 & influenza a/b nucleic acid test on the cobas® liat® system.The alleged sample initially generated a positive result for all targets; sars-cov-2, influenza a, and influenza b.The same sample was retested on a different platform (cepheid) which yielded a negative result for all targets.The patient also tested negative with an antigen test after quarantining.The initial positive results were released.No harm was alleged.An investigation was conducted to evaluate the customer issue.Per fda¿s eua guidance, 1 mdr will be filed.
 
Manufacturer Narrative
An investigation was conducted to evaluate the customer issue.Note, while the customer indicated the alleged sample was tested on (b)(6) 2022, the provided data indicated the sample was tested on (b)(6) 2022.Throughout the data analysis, a systematic issue was not observed and a product problem was not found.Overall, the most likely cause for the discrepant results may be due to a reagent or sample handling issue or cross-contamination.In addition, the lod and the type of influenza viral strains detected differ significantly between the cobas® sars-cov-2 & influenza a/b assay on the cobas® liat system and the cepheid genexpert assays.As such, results with one assay may not be reproducible with a different assay.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS® SARS-COV-2 & INFLUENZA A/B
Type of Device
COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES
Manufacturer (Section D)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer (Section G)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
na
branchburg NJ 08876
Manufacturer Contact
stacie-ann creighton
1080 us hwy 202 s
na
branchburg, NJ 08876
9082537112
MDR Report Key14402889
MDR Text Key300163819
Report Number2243471-2022-00473
Device Sequence Number1
Product Code QLT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EUA201779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Catalogue Number09211101190
Device Lot Number20131L
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2022
Initial Date FDA Received05/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexFemale
-
-