• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CODMAN 11MM DISPOSBL PERFORATR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CODMAN 11MM DISPOSBL PERFORATR; DISPOSABLE PERFORATORS Back to Search Results
Model Number 261222
Device Problem Melted (1385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the top of the green plastic of a perforator melted and was producing smoke while in use.The perforator was inspected prior to use as instructed; the spring tests were performed; the plastic sleeve was checked prior to use.No patient injury reported and the event did not led to surgical delay.
 
Manufacturer Narrative
Perforator was not returned for evaluation (discarded); therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN 11MM DISPOSBL PERFORATR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14414923
MDR Text Key291819509
Report Number3014334038-2022-00098
Device Sequence Number1
Product Code HBF
UDI-Device Identifier10381780513605
UDI-Public10381780513605
Combination Product (y/n)N
PMA/PMN Number
K071931
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number261222
Device Catalogue Number261222
Device Lot NumberJ83A95
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/29/2022
Date Device Manufactured06/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-