COOK INC TUOHY-BORST LARGE BORE CLEAR PLASTIC SIDEARM ADAPTER; KGZ ACCESSORIES, CATHETER Back to Search Results
Catalog Number PTBYC-RA
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Per a search of sales data, possible lot numbers are 10069195 and 10262383.Pma/510(k) number = exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
As reported, during a lower extremity angiogram, blood leaked from a tuohy-borst large bore clear plastic sidearm adapter.While injecting contrast through the device, a back-flow of blood was noted from the wire port.Other manufacturers' 0.014-inch and 0.018-inch wires were used through the device.The complaint device was used to complete the procedure.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event Description
No additional information has been received since the last report was submitted.
 
Manufacturer Narrative
Summary of event: as reported, during a lower extremity angiogram, blood leaked from a tuohy-borst large bore clear plastic sidearm adapter.While injecting contrast through the device, a back-flow of blood was noted from the wire port.Other manufacturers' 0.014-inch and 0.018-inch wires were used through the device.The complaint device was used to complete the procedure.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record, manufacturing instructions, and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation.A document-based investigation evaluation was performed.A search of sales history revealed two possible lots associated with this complaint (10069195 or 10262383).No related non-conformances were found on either possible lot, and there have been no other reported complaints for either lot number.The information provided upon review of the dmr and dhr suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that a component failure unrelated to manufacturing or design deficiencies contributed to this incident.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUOHY-BORST LARGE BORE CLEAR PLASTIC SIDEARM ADAPTER
Type of Device
KGZ ACCESSORIES, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14431121
MDR Text Key300255891
Report Number1820334-2022-00734
Device Sequence Number1
Product Code KGZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPTBYC-RA
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/05/2022
Initial Date FDA Received05/18/2022
Supplement Dates Manufacturer Received07/29/2022
Supplement Dates FDA Received08/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment