Model Number ROB10013 |
Device Problems
Unstable (1667); Connection Problem (2900); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that during a cori assisted tka surgery the real intelligence robotic drill had system time out error.They quit the case, created a new case and tried again but again they got the system time out error.They exited out of the case and the system came up with a robotic drill critical error warning.They went to run a drill diagnostic test and when they attempted to insert the bur, they found it would not lock into the drill.They tried to swap out the ri robotic drill attachment, but it would not unlock from the real intelligence robotic drill.The procedure was completed, with a significant delay, using a s+n back-up device.Patient was not harmed beyond the problem reported.After the case, they were able to get the attachment off the drill, but the drill still failed all the checks on the drill diagnostic test and a bur would not lock in.
|
|
Manufacturer Narrative
|
Section h10: the cori drill, pn (b)(4) , sn (b)(6) , use in treatment, was returned for evaluation.A relationship between the reported event and the device could not be established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was not confirm.The drill was tested using a tester long attachment.The drill passed kpc test and a case was review successfully without any error message or inconvenience removing the tester long attachment.The bur was load and secured during evaluation.The drill was found to operated as design.Although the reported problem was not confirmed, factor that may have contributed to the reported symptom may have been associated with an intermittent failure of the exposure motor.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of this case.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored through complaint investigation and post market surveillance activities.Internal complaint reference number: case (b)(4) section d10 was corrected.
|
|
Manufacturer Narrative
|
H3, h6: the cori drill, pn rob10013, (b)(6) , use in treatment, was returned for evaluation.A relationship between the reported event and the device could not be established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was not confirm.The drill was tested using a tester long attachment.The drill passed kpc test and a case was review successfully without any error message or inconvenience removing the tester long attachment.The bur was load and secured during evaluation.The drill was found to operated as design.Although the reported problem was not confirmed through a visual or functional evaluation, factors that may have contributed to the reported symptom may be associated with an intermittent failure of the drill exposure motor due to the thermo-mechanical stress induced within the motor at the encoder and electrical noise on the console error status inputs to the drill exposure motor encoder.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.A historical escalation event review was completed.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A historical review concluded that the lot, serial number or part number reported in this event is related to a corrective/preventive action already implemented.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Continuous improvements have been made to the cori robotic drill and manufacturing processes to reduce drill disconnection error messages.These improvements consisted of: 1.A hardware update to the cori console to reduce noise on the internal electronics.2.An update to the cori system¿s software and firmware to improve the user experience when error messages are displayed, and 3.A hardware update to the cori drill to reduce mechanical stress on drill exposure the motor.The first two improvements are fully deployed.The third improvement is being deployed for new orders and as drills are returned for routine servicing.Also, smith+nephew is voluntarily performing a recall/field notification for the cori real intelligence robotic drill.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|
|
Search Alerts/Recalls
|
|