• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Model Number SGC0701
Device Problems Break (1069); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2022
Event Type  malfunction  
Event Description
This is filed to report the steerable guide catheter cable break.It was reported that during preparation of the steerable guide catheter (sgc), a pop sound was heard when adding negative on the +/-knob.A cable break is suspected.The sgc was not used.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information and without the device to analyze, a cause of the speculated cable break could not be determined.The reported popping noise appears to be a result of the cable break.There is no indication of a product quality issue with respect to manufacture, design or labeling.The additional mitraclip device referenced is filed under a separate medwatch report number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14432534
MDR Text Key292047232
Report Number2024168-2022-05339
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2022
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number11130R303
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexMale
Patient Weight54 KG
-
-