The review of the device history record showed no deviations.The product complies with the specifications valid at the time of manufacture.The complained polyethylene ps plateau was returned for investigation.The visual inspection of the plateau revealed wear marks and deformations at the distal surface, medial/lateral snap, central rail and the anterior edge.The medial/lateral locking mechanism shows marks of the tibial tray and was bent proximally.It is assumed that the anterior locking mechanism was not engaged correctly leaving the plateau lie obliquely on the tibial tray after the surgery was finished.The plateau stayed inside the body for 5 weeks, as a result bent under constant load leading to deformation and wear effects.Based on the surgical technique, it is important to ensure the correct assembly of the plateau into the tibial tray before impaction.One of the reason could be assumed that the tibial or the femoral rotation may have caused the post/box mechanism and/or the condyles to guide the snap mechanism incorrectly.As a measure, an intensive exchange of training material is underway with the users in order to clarify the correct assembly of the pe plateau to the tibial component.Additionally, an engineering change request is initiated to provide a more detailed surgical technique.The dimensions of the entire lot were measured 100% during production and no deviations were found.Based on the above information no product problem is assumed.
|