• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL X1 CURVED 37CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL X1 CURVED 37CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSLX137C
Device Problem Difficult or Delayed Activation (2577)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/10/2021
Event Type  malfunction  
Event Description
It was reported that during thoracoscopy esophagectomy, partial gastrectomy by videolaparoscopy and cervical esophagogastroplasty, the enseal tweezers presented a defect and ceased to function in the middle of the procedures (it did not seal and warned that it had not completed the coagulation process), requiring the use of an ultrasonic harmonic tweezers to complete the procedures.
 
Manufacturer Narrative
(b)(4).Batch #: u95615.Additional information was requested and the following was obtained: has the patient suffered any adverse consequences? no investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the nslx137c device was returned with the jaws misaligned.Upon visual inspection to the jaws it was observed that one side of the hinge pin was missing.The hinge pin missing was not returned with the device.The device was tested on the generator, this resulted in the ¿reposition jaws and reactivate¿ alert screen, this is advising the generator cannot deliver energy.Then the "replace instrument" screen would be displayed after receiving the "reposition jaws and reactivate" message three time.The hinge pin missing could have contributed to the reported ¿tissue effect issues¿ , this due to the reduced compression capacity of the jaws.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery¿s quality system.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances related to the reported complaint condition were identified.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL X1 CURVED 37CM SHAFT
Type of Device
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
MDR Report Key14434524
MDR Text Key292206386
Report Number3005075853-2022-03243
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10705036015468
UDI-Public10705036015468
Combination Product (y/n)N
PMA/PMN Number
K172580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNSLX137C
Device Catalogue NumberNSLX137C
Device Lot NumberU95615
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2022
Date Manufacturer Received04/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GENERATOR
-
-