• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE); PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE); PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number UNKNOWN
Device Problems Malposition of Device (2616); Biocompatibility (2886)
Patient Problems Infiltration into Tissue (1931); Inflammation (1932); Pain (1994); Deformity/ Disfigurement (2360); Metal Related Pathology (4530)
Event Date 01/01/2016
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).Samuel, l.T., zuke, w.A., mahmood, s., munim, m.A., alamir, p., & brooks, p.J.(2022).Hip resurfacing: a single surgeon us series with minimum ten-year follow-up.The journal of arthroplasty.Doi: https://doi.Org/10.1016/j.Arth.2022.04.008.
 
Event Description
It was reported that on literature review "hip resurfacing: a single surgeon u.S.Series with minimum ten-year follow-up¿ 8.3 years after a bhr surgery, one (1) patient had pain, high levels of metal ions (56.7 ppb of co and 29.8 ppb of cr), cup malposition with excessive anteversion and effusion.Revision surgery was performed to explant both implants.Competitor devices were implanted.The patient outcome and current health status is unknown.No further information is available.
 
Manufacturer Narrative
Additional information: a2, a3, h6 (component code, type of investigation, investigation findings, investigation conclusions).Results of investigation: it was reported that the patient was revised due to pain, high levels of metal ions, cup malposition with excessive anteversion and effusion.The devices, used in treatment, were not returned for analysis.As no device part and batch numbers were provided for investigation, a complaint history review, manufacturing record review, device labelling / ifu, or escalation actions review could not be performed.If more information is received, this investigation will be reopened.A risk management review was performed for the devices involved.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.Without clinically relevant patient-specific supporting documentation, a thorough medical investigation cannot be rendered nor can the root cause of the reported failure be determined.However, we cannot rule out the reported cup malposition with excessive anteversion as a likely contributory factor.The impact to the patient beyond that which has already been reported cannot be confirmed nor concluded.Therefore, no further medical assessment is warranted at this time.Should any additional relevant clinical information be provided, this complaint would be re-assessed.Based on the information provided, including elevated ion levels, we can confirm the reported complaint; however our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.Corrected data: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE)
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14438957
MDR Text Key292062778
Report Number3005975929-2022-00291
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Literature,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/22/2022
Initial Date FDA Received05/19/2022
Supplement Dates Manufacturer Received09/12/2022
Supplement Dates FDA Received09/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age47 YR
Patient SexFemale
-
-